Applied Silicone Corporation PN 40096 Medical Grade Primer for Metals and Plastics contains 10% moisture-reactive silanes in xylene that improve adhesion of silicone elastomers to some substrates such as metals, ceramics and plastics.

This product is well-suited for coupling Applied Liquid Silicone Rubbers, High Consistency Rubbers, and Adhesives to metals, fabrics, plastics, ceramics, and other substrates. Special features include:

•  Water consistency for easy handling 
•  Applicable by spraying, dipping, or brushing  
•  Requires no heating 
•  Improves silicone elastomer bonds to most substrates 


BIOCOMPATIBILITY 
 
This material is designed to pass ISO 10993 requirements for long term implantation when encapsulated in an ASC Implant Grade Silicone Elastomer.  Additional testing is available upon request. 
 

APPLICATIONS 

The surface to be primed should be thoroughly cleaned and allowed to dry. The primer is then applied by brushing, dipping, wiping, or spraying. A thin film will give the best results. Thickness can be controlled by thinning the solution in aromatic or aliphatic organic solvents. The primed substrate must be exposed to atmospheric moisture immediately after application for thirty to sixty minutes to induce evaporation of the solvent and to cure  the silane coupling agents.  Prolonged exposure greater than two hours will reduce effectiveness. A minimum of thirty percenty atmospheric relative humidity is recommended for full cure, and therefore, maximum adhesion to substrate. Air exposure of the primer should occur immediately after application to the clean surface to be primed. If immediate exposure to air containing at least thirty percent relative humidity does not occur, the resulting primed film may result in poor adhesion to both the substrate and to subsequent silicone material added to the primed substrate. 
 


PRECAUTIONS 

A Material Safety Data Sheet (MSDS) is supplied with each shipment. The primer hydrolyzes quickly upon contact with atmospheric moisture.  Keep in tightly sealed containers and avoid exposing surface of unused material to ambient air. Purging containers with dry nitrogen before resealing is recommended. The dispersion contains the solvent xylene which is a flammable liquid and a health hazard. The dispersion should not come into contact with eyes, skin, or clothing.  Do not breathe the vapor.  Use the dispersion with adequate ventilation. Keep away from open flame, sources of sparks, and heat. When transferring dispersion, electrically ground all vessels. The user is urged to be familiar with the properties of xylene.
 


PACKAGING AND STORAGE 

Packaging is available in 1-pint, 1-gallon and 5-gallon cans and in 55-gallon drums. 

The solution is a flammable liquid and must be stored and transported in accordance with applicable Federal, State, and Local regulations. Store in the original sealed container below 30ºC.

Suggested retest date is twelve months from date of shipment. 


LOT TESTING 

Each production lot is certified as having specified physical properties. Other testing is available upon request. 
 


TYPICAL PROPERTIES These values are typical and are not intended for use in preparing specifications.
 
TEST RESULTS
PN 40096
 
COLOR CLEAR
SPECIFIC GRAVITY @ 25ºC 0.86 - 0.88
PERCENT SOLIDS 2
FLASH POINT 80ºF
 

DISCLAIMER 
 

Applied Silicone Corporation warrants that this product meets applicable descriptions and specifications for a period of six months from shipment. No other warranties, expressed or implied, are intended. This product was designed for sophisticated users. The user must carefully review literature for uses where adverse effects are reported for certain applications of silicone in medical devices. The user of this material must independently determine that any product incorporating this material is safe and effective. The user of this material must thoroughly test any application consistent with industry requirements and/or standards.  Applied Silicone Corporation disclaims any expressed or implied warranty against the infringement of any patent. Any suggestion of use by Applied Silicone should not be taken 
as inducement to infringe on any particular patent. 

Revised 08.19.02

270 Quail Court 
Santa Paula, California 93060
Phone: 805-525-5657 

Fax: 805-933-1675
Internet: www.appliedsilicone.com 
E-mail:  siliconesales@appliedsilicone.com


 

©2002ASC          July 1998-02TB

Home Products MSDS Contact Us Legal