DESCRIPTION Applied Silicone Corporation PN 40096 Medical Grade Primer for Metals and Plastics contains 10% moisture-reactive silanes in xylene that improve adhesion of silicone elastomers to some substrates such as metals, ceramics and plastics. This product is well-suited for coupling Applied Liquid Silicone Rubbers, High Consistency Rubbers, and Adhesives to metals, fabrics, plastics, ceramics, and other substrates. Special features include: • Water consistency
for easy handling
|
||||||||||||||
|
BIOCOMPATIBILITY This material is designed to pass ISO 10993 requirements for long term implantation when encapsulated in an ASC Implant Grade Silicone Elastomer. Additional testing is available upon request. |
||||||||||||||
|
APPLICATIONS The surface to be primed
should be thoroughly cleaned and allowed to dry. The primer is then applied
by brushing, dipping, wiping, or spraying. A thin film will give the best
results. Thickness can be controlled by thinning the solution in aromatic
or aliphatic organic solvents. The primed substrate must be exposed to
atmospheric moisture immediately after application for thirty to sixty minutes to
induce evaporation of the solvent and to cure the silane coupling
agents. Prolonged exposure greater than two hours will reduce effectiveness.
A minimum of thirty percenty atmospheric relative humidity is recommended for full
cure, and therefore, maximum adhesion to substrate. Air exposure of the
primer should occur immediately after application to the clean surface
to be primed. If immediate exposure to air containing at least thirty percent relative
humidity does not occur, the resulting primed film may result in poor adhesion
to both the substrate and to subsequent silicone material added to the
primed substrate.
|
||||||||||||||
|
PRECAUTIONS A Material Safety Data Sheet
(MSDS) is supplied with each shipment. The primer hydrolyzes quickly
upon contact with atmospheric moisture. Keep in tightly sealed
containers and avoid exposing surface of unused material to ambient air.
Purging containers with dry nitrogen before resealing is recommended. The
dispersion contains the solvent xylene which is a flammable liquid and
a health hazard. The dispersion should not come into contact with eyes,
skin, or clothing. Do not breathe the vapor. Use the dispersion
with adequate ventilation. Keep away from open flame, sources of sparks,
and heat. When transferring dispersion, electrically ground all vessels.
The user is urged to be familiar with the properties of xylene.
|
||||||||||||||
|
PACKAGING AND STORAGE Packaging is available in 1-pint, 1-gallon and 5-gallon cans and in 55-gallon drums. The solution is a flammable liquid and must be stored and transported in accordance with applicable Federal, State, and Local regulations. Store in the original sealed container below 30ºC. Suggested retest date is twelve months from date of shipment. |
||||||||||||||
|
LOT TESTING Each production lot is certified
as having specified physical properties. Other testing is available upon
request.
|
||||||||||||||
|
||||||||||||||
|
DISCLAIMER Applied Silicone Corporation
warrants that this product meets applicable descriptions and specifications
for a period of six months from shipment. No other warranties, expressed
or implied, are intended. This product was designed for sophisticated users.
The user must carefully review literature for uses where adverse effects
are reported for certain applications of silicone in medical devices. The
user of this material must independently determine that any product incorporating
this material is safe and effective. The user of this material must thoroughly
test any application consistent with industry requirements and/or standards.
Applied Silicone Corporation disclaims any expressed or implied warranty
against the infringement of any patent. Any suggestion of use by Applied
Silicone should not be taken
Revised 08.19.02 |
||||||||||||||
|