PRODUCT INFORMATION SHEET

APPLIED SILICONE CORPORATION

CUSTOM HIGH CONSISTENCY PEROXIDE CURE SILICONE (HCRP 50X) MASTERBATCH


DESCRIPTION 
 
Applied’s Custom High Consistency Peroxide Cure Silicone (HCRP 50X) Masterbatch is a one part, silicone elastomer designed to blend into 40045 HCRP or other HCRP products to improve adhesion. 

HCRP Masterbatch contains 10 pph 2,4-dichlorobenzoyl peroxide catalyst which, when blended between 0 pph - 5 pph, will initiate cure between 110°C and 175°C. Postcure is required to remove peroxide residues.   


BIOCOMPATIBILITY 
 
When blended into ASC 40045 HCRP-50 at <10%, the resulting material was designed to pass ISO 10993 requirements for long term implantation. 

APPLICATIONS 
 
This product was designed to be added to suitable HCRP bases between 0 pph to 5 pph to improve silicone to silicone adhesion in compression and transfer molding. Cure is typically achieved in sixty seconds at 165ºC. Postcure is necessary to stabilize physical properties and remove peroxide residues. Typical post-cure for a 4 mm thick sample is one hour at 200ºC or two hours at 170ºC. 

PRECAUTIONS 

A material safety data sheet is supplied with the shipment. Contact with eyes may cause temporary discomfort. Avoid contact with eyes.

Material should be two roll mill freshened before use. Avoid overheating when refreshing because this may cause premature vulcanization.  


PACKAGING AND STORAGE 
 
This product is available in 50 pound boxes.  

This product should be stored under cool dry conditions below 30ºC. Containers should be sealed to avoid contamination.  

Suggested retest at six months from the date of shipment. 


TYPICAL PROPERTIES 

When blended with 0 pph - 5 pph masterbatch to 100 pph 40045 HCRP-50; press cured three minutes @ 120°C and postcured one hour @ 200°C.

These values are typical and are not intended for use in preparing specifications.
TEST RESULTS
DUROMETER, SHORE A 50
TENSILE STRENGTH, psi 1400
ELONGATION, % 600
TEAR STRENGTH, ppi 190
SPECIFIC GRAVITY 1.15
APPEARANCE TRANSLUCENT, GUM
CONSISTENCY GUM

DISCLAIMER 
 
Applied Silicone Corporation warrants that this product meets applicable descriptions and specifications for a period of six months from shipment.  No other warranties, expressed or implied, are intended.  This product was designed for sophisticated users.  The user must carefully review literature for uses where adverse effects are reported for certain applications of silicone in medical devices.  The user of this material must independently determine that any product incorporating this material is safe and effective.  The user of this material must thoroughly test any application consistent with industry requirements and/or standards.  Applied Silicone Corporation disclaims any expressed or implied warranty against the infringement of any patent.  Any suggestion of use by Applied Silicone should not be taken as inducement to infringe on any particular patent. 
 